Smoke box inside size |
Long 914mm±3mm, Wide 610mm±3mm, high 914mm±3mm. |
Instrument size |
Long 1110mm±10mm, Wide 750mm±10mm, high 2040mm±10mm. |
Power |
220v,4kw. (The total power of the device should be indicated) |
Measuring range |
0.0001-100%; |
Measuring accuracy |
±3%; |
Working voltage |
200-240v 50Hz; |
Ambient temperature |
room temperature-40 ℃ |
Relative temperature |
≤85%; |
Working environment |
Instrument operation, should avoid strong light direct, and no forced air flow phenomenon. |
Measuring software |
its Windows XP operating interface, the main data can be tested smoke light plot, specific light density, quality light density, quality loss, quality loss rate, the average smoke density. Also has the calibration, the editing, the test, the parameter adjustment and so on Operation interface. |
Sample size |
75x75mm. |
Light source voltage |
6.5V |
operating voltage |
4v±0.2v. |
Light temperature |
2200k±100k. |
Filter |
Enlarge the optical density of the filter 2.0; The optical density of the filter is 3.0, and the optical density of the filter is 0.5, one is 3. |
Heating cone |
Power 2600W, |
temperature control range |
0-1000 ℃, thermocouple temperature, equipped with automatic cold end compensation. |
Heat Flow radiometer |
Heat flow meter signal computer acquisition calculation, equipped with water-cooled piping. |
Auxiliary heater |
Power 500W, can make the test chamber internal temperature stability in 40℃±10℃ or 55℃±10℃. |
Gas |
More than 95% purity of the 3.5kpa±kpa, the minimum pressure for the. |
Weight |
About 500Kg. |
Instrument Size |
1.22mx0.68mx2m |
The configuration standard |
conforms to the ISO 5659-2 2006,GBt8323.2-2008 |
Desktop (including display) |
one |
Opaque template one block |
one |
Spare bulbs |
two |
Manual |
one |
Certificate |
one |
Calibration lenses |
two pieces |
Printer |
one |
Analysis software |
one |
Movable light gate |
one |
Blasting aluminum foil |
one |
Stainless Steel Sample box |
two |